Multicenter evaluation of an intrapericardial left ventricular assist system

Abstract

Objectives: The aim of this study was to conduct an initial clinical evaluation of the new HeartWare Ventricular Assist System (HeartWare, Inc., Framingham, Massachusetts) in a multicenter, prospective, nonrandomized single-arm clinical trial.

Background: Heart failure is a worldwide epidemic. The effectiveness of heart transplantation and medical therapy is limited, resulting in the emergence of mechanical circulatory support as a primary treatment for end-stage heart disease. Left ventricular assist devices that use rotary pumps are small and durable, which might reduce morbidity and mortality during support.

Methods: Fifty heart transplant candidates with New York Heart Association functional class IV symptoms were supported at 5 international centers by the HeartWare System for 180 days, until heart transplant, myocardial recovery and device explant, or death. Patients who continue to be supported have been followed for a minimum of 2 years.

Results: Of the 50 patients, 20 (40%) received transplants, 4 (8%) had the pump explanted after myocardial recovery, and 17 (34%) continue support at 2 years. Nine (18%) patients died during support from sepsis (n = 3), multiple organ failure (n = 3), or hemorrhagic stroke (n = 3). The actual survival at 6, 12, and 24 months was 90%, 84%, and 79%, respectively. In the survivors, measures of quality of life showed a significant improvement over baseline values. Significant improvements were found for recognition memory at 3 months after implant (p = 0.006). The most frequent adverse events were infection and bleeding.

Conclusions: Patients with end-stage heart failure can be safely and effectively supported by the HeartWare Ventricular Assist System with improved quality of life and neurocognitive function.